![background image](http://html.mh-extra.com/html/keithley/6430/6430_instruction-manual_661765410.webp)
17-100 SCPI Command Reference
:POINts <n>
:TRACe:POINts <n>
Specify buffer size
Parameters
<n> =
1 to 2500
Specify buffer size
MINimum
1
MAXimum
2500
DEFault
100
Query
:POINts?
Query buffer size
:POINts? MINimum
Query smallest allowable buffer size
:POINts? MAXimum
Query largest allowable buffer size
:POINts? DEFault
Query *RST default buffer size
Description
This command is used to specify the size of the buffer.
:ACTual?
:TRACe:POINts:ACTual?
Query number of stored readings
Description
This query command is used to determine how many stored readings are in
the buffer. After sending this command and addressing the unit to talk, the
number of readings stored in the buffer will be sent to the computer.
:FEED <name>
:TRACe:FEED <name>
Specify readings source
Parameters
<name> =
SENSe[1]
Put raw readings in buffer
CALCulate[1]
Put Calc1 readings in buffer
CALCulate2
Put Calc2 readings in buffer
Query
:FEED?
Query buffer feed
Description
This command is used to select the source of readings to be placed in the
buffer. With SENSe[1] selected, raw readings are placed in the buffer when
storage is performed.
With CALCulate[1] selected, math expression results (Calc1) are placed in
the buffer. With CALCulate2 selected, Calc2 readings are placed in the
buffer.
TRACe:FEED cannot be changed while buffer storage is active. See
Section 8 for more information.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......