Sweep Operation
Step 5: Run sweep.
To run the sweep, press the SWEEP key. After the sweep is completed, turn the output off by
pressing the ON/OFF OUTPUT key.
Step 6: Read buffer.
Use the RECALL key to access the source-measure readings stored in the buffer. Use the
TOGGLE to display statistical information.
Performing a custom sweep
Step 1: Configure source-measure functions.
Configure the SourceMeter for the desired source-measure operations as follows:
1.
Select the desired source function by pressing SOURCE V or SOURCE I.
2.
Set the source level and compliance limit to the desired values.
3.
Press MEAS V or MEAS I to select the desired measurement function, then choose the
desired measurement range.
Step 2: Configure sweep.
Configure the sweep as follows:
1.
Press CONFIG then SWEEP.
2.
Select TYPE, then press ENTER.
3.
Select CUSTOM, then press ENTER.
4.
Use the displayed menu selections to enter the desired # POINTS, individual point val-
ues (ADJUST POINTS), and INIT (initial) value.
5.
From the CONFIGURE SWEEPS menu, select SWEEP COUNT, press ENTER, then
choose FINITE or INFINITE as desired.
6.
Again from the CONFIGURE SWEEPS menu, choose SOURCE RANGING, press
ENTER, then select BEST FIXED, AUTO RANGE, or FIXED as appropriate.
7.
Press EXIT to return to normal display.
Step 3: Set delay.
Set the source delay as follows:
1.
Press CONFIG then SOURCE V or SOURCE I depending on the selected source
function.
2.
Select DELAY, then press ENTER.
3.
Set the delay to the desired value, then press ENTER.
4.
Press EXIT to return to normal display.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......