![background image](http://html.mh-extra.com/html/keithley/6430/6430_instruction-manual_661765493.webp)
Status and Error Messages
B-5
Number
Error message
Event
1
Status register
2
Bit
-105
-104
-103
-102
-101
GET not allowed
Data type error
Invalid separator
Syntax error
Invalid character
EE
EE
EE
EE
EE
Standard Event
Standard Event
Standard Event
Standard Event
Standard Event
5
5
5
5
5
-100
Command error
EE
Standard Event
5
+000
No error
SE
+100
+101
+102
+103
+104
+105
+106
+107
+108
+109
+111
+112
+113
+114
Measurement events:
Limit 1 failed
Low limit 2 failed
High limit 2 failed
Low limit 3 failed
High limit 3 failed
Active limit tests passed
Reading available
Reading over
fl
ow
Buffer available
Buffer full
OUTPUT interlock asserted
Temperature limit exceeded
Voltage limit exceeded
Source in compliance
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
0
1
2
3
4
5
6
7
8
9
11
12
13
14
+200
+300
+303
Standard events:
Operation complete
Operation events:
Device calibrating
Device sweeping
SE
SE
SE
Standard Event
Operation Event
Operation Event
0
0
3
1
EE = Error Event
SE = Status Event
SYS = System Error Event
2
Use following queries to read status registers:
Standard Event: *ESR?
Operation Event: STAT:OPER?
Measurement Event: STAT:MEAS?
Questionable Event: STAT:QUES?
Table B-1 (cont.)
Status and error messages
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......