10-10
Triggering
The Trigger Link connections for this test system are shown in Figure 10-6. Trigger Link of
the SourceMeter is connected to Trigger Link (IN or OUT) of the switching mainframe. Note
that with the default trigger settings of the switching mainframe, line #1 is an input, and line #2
is an output.
For this example, the SourceMeter and Switching Mainframe are configured as follows.
SourceMeter setup
Step 1: Restore bench defaults
Press the MENU key, select SAVESETUP, then press the ENTER key. From the
SAVESETUP menu, select GLOBAL, then press ENTER. From the GLOBAL SETUP
MENU, select RESET, then press ENTER. Select BENCH, then press ENTER.
Step 2: Set up trigger parameters
Press the CONFIG key, and then the TRIG key to access the CONFIGURE TRIGGER
menu. Select TRIG-LAYER, and then press ENTER to access the CONFIGURE TRIG-
LAYER menu.
Step 3: Set trigger-in event to TRIGGER-LINK
Select TRIGGER-IN, then press ENTER. Select TRIGGER-LINK, then press ENTER.
Step 4: Set trigger input line to #2
Select #2, then press ENTER three times to return to CONFIGURE TRIG-LAYER menu.
Step 5: Set trigger output line to # 1
Select TRIGGER-OUT, then press ENTER. Select LINE, then press ENTER. Select #1,
then press ENTER.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN USA
7001 or 7002 Switch System
6430 SourceMeter
Trigger
Link
Trigger
Link Cable
(8501)
Trigger
Link
OUT
IN
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V,
Ω
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Figure 10-6
Trigger link
connections
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......