19-10
Calibration
Step 3: Current calibration
Perform the steps below for each current range using Table 19-4 as a guide.
1.
Connect the Model 6430 to the digital multimeter as shown in Figure 19-2. (Connect
Model 6430 INPUT/OUTPUT HI to the DMM AMPS input; INPUT/OUTPUT LO to
DMM INPUT LO.) Select the multimeter DC current measurement function.
2.
From normal display, press the SOURCE I key.
3.
Press the EDIT key to select the source display
fi
eld, and then use the down RANGE
key to select the 1µA source range.
4.
From normal display, press MENU.
5.
Select CAL, and then press ENTER. The unit will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
SAVE LOCK CHANGE-PASSWORD
6.
Select EXECUTE, and then press ENTER. The instrument will display the following
message:
I-CAL
Press ENTER to 1.0000µA
7.
Press ENTER. The Model 6430 will 1µA and simultaneously display the
following:
DMM RDG: +1.000000µA
Use
,
▲
,
▼
,
, ENTER or EXIT.
8.
Note and record the DMM reading, and then adjust the Model 6430 display to agree
exactly with the actual DMM reading. Use the up and down arrow keys to select the
digit value, and use the left and right arrow keys to choose the digit position (or use the
Model 6430
Digital Multimeter
Input LO
Amps
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V,
Ω
,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Figure 19-2
Mainframe current
calibration
connections
▲
▲
▲
▲
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......