:POINts <n>
:SOURce[1]:SWEep:POINts <n>
Set source-measure points for sweep
Parameters
<n> =
1 to 2500
Specify number of source-measure points
MINimum
1
MAXimum
2500
DEFault
2500
Query
:POINts?
Query number of sweep points
:POINts? DEFault
Query *RST default number of sweep points
:POINts? MINimum
Query lowest allowable number of sweep
points
:POINts? MAXimum
Query highest allowable number of sweep
points
Description
The :POINts command speci
fi
es the total number of source-measure points
in a sweep. For a linear sweep, the source-measure points are equally spaced
(stepped) between the start level and the stop level. For a log sweep, the
source-measure points are equally spaced on a logarithmic scale. Note that
the start and stop levels are source-measure points.
Step size for a linear sweep can be calculated as follows:
Step Size = (Stop - Start) / (Points - 1)
Step Size = Span / (Points -1)
Step size for a logarithmic sweep can be calculated as follows:
An alternate way to set the source-measure points in a sweep is to specify
the step size using the :STEP command.
Note that the :POINts and :STEP commands are coupled. Changing the
number of source-measure points also changes the step size. Conversely,
changing the step size changes the number of source-measure points.
:DIRection <name>
:SOURce[1]:SWEep:DIRection <name>
Set direction of sweep
Parameters
<name> =
UP
Run sweep from start to stop
DOWn
Run sweep from stop to start
Query
:DIRection?
Query direction of sweep
Description
Normally, a sweep is run from the start level to the stop level. The :STARt
and :STOP, or :CENTer and :SPAN commands are used to set these levels.
This command lets you change the execution direction of the sweep. With
DOWn selected, the sweep will begin at the stop level and end at the start
level. Selecting UP restores sweep operation to the normal start to stop
direction.
Log Step Size
log 10 (Stop)- log10 (Start)
Points - 1
-----------------------------------------------------------------
=
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......