Remote Operations
13-7
REN (remote enable)
The remote enable command is sent to the SourceMeter by the controller to set up the instru-
ment for remote operation. Generally, the instrument should be placed in the remote mode
before you attempt to program it over the bus. Setting REN true does not place the instrument
in the remote state. You must address the instrument to listen after setting REN true before it
goes into remote.
The SourceMeter must be in remote in order to use the following commands to trigger and
acquire readings:
•
:INITiate and then :FETCh?
•
:READ?
•
:MEASure?
IFC (interface clear)
The IFC command is sent by the controller to place the SourceMeter in the local, talker, lis-
tener idle states. The unit responds to the IFC command by cancelling front panel TALK or
LSTN lights, if the instrument was previously placed in one of these states.
Note that this command does not affect the status of the instrument. Settings, data, and event
registers are not changed.
With auto output off enabled (:SOURce1:CLEar:AUTO ON), the output will remain on if
operation is terminated before the output has a chance to automatically turn off.
To send the IFC command, the controller need only set the IFC line true for a minimum of
100µs.
LLO (local lockout)
Use the LLO command to prevent local operation of the instrument. After the unit receives
LLO, all of its front panel controls except OUTPUT OFF are inoperative. In this state, pressing
LOCAL will not restore control to the front panel. The GTL command restores control to the
front panel. Cycling power will also cancel local lockout.
GTL (go to local)
Use the GTL command to put a remote-mode instrument into local mode. The GTL com-
mand also restores front panel key operation.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......