Connections
2-11
The basic connection scheme for guarded ohms measurements is shown in Figure 2-5. If the
guard resistance path (R
G
) is <1k
Ω
, IR drop in the GUARD test lead could be high enough that
the guard voltage at the resistor network is significantly less than the output voltage at the DUT.
This results in leakage current and will corrupt the measurement. To cancel the effect of IR
drop in the GUARD test lead, connect GUARD SENSE to the resistor network as shown in
Figure 2-6. Guard sense ensures that the guard voltage at the resistor network is the same as the
output voltage at the DUT.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V,
Ω
,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Triax Cable
HI
DUT
WARNING Guard is at the same voltage level as
input/output high. Therefore, if a
hazardous voltage (
≥
42V peak) is
present on input/output high, it is
also present on the guard terminals
of the Remote PreAmp and mainframe.
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
LO
Banana Plug Cables
6430 Mainframe (rear panel connectors)
Resistor
Network
Ohms
Guard
I
G
<1k
Ω
Guard
Sense
IN/OUT
HIGH
Preamp
Cable*
R
G
*Keithley part number: CA-176-1D.
Figure 2-6
Guarded ohms
connections
(guard sense)
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......