ADDITIONAL SOURCE SPECIFICATIONS
COMMAND PROCESSING TIME:
Maximum time required for the output to begin to change following the receipt of :SOURce:VOLTage|CURRent
<nrf> command.
Autorange On:
10ms.
Autorange Off:
7ms.
OUTPUT SETTLING TIME (typical to 10% of final value):
<2s, 1pA and 10pA ranges; <50ms, 100pA through 10nA ranges; <5ms, 100nA through
100mA ranges.
OUTPUT SLEW RATE:
30V/ms, any V range, 10mA compliance.
COMMON MODE VOLTAGE:
±42VDC maximum.
4-WIRE SENSE:
Up to 1V drop per load lead, 10
Ω
maximum per sense lead, 100µA range and up. For details on using 4-wire sense with the 10µA
range and below, refer to the User
’
s Manual.
OVER TEMPERATURE PROTECTION:
Internally sensed temperature overload puts unit in standby mode.
RANGE CHANGE OVERSHOOT:
Overshoot into a fully resistive 100k
Ω
load, 10Hz to 1MHz BW, adjacent ranges, 100mV typical, except 20V/200V
range boundary.
MINIMUM COMPLIANCE VALUE:
0.1% of range.
MEASURE SPECIFICATIONS
1
Voltage Measurement Accuracy (4-wire sense)
3
Max.
Input
2
Accuracy (23
°
C ± 5
°
C)
Range
Resolution
Resistance
1 Year, ±(%rdg + volts)
200.000 mV
1 µV
>10
16
Ω
0.012% + 350 µV
2.00000
V
10 µV
>10
16
Ω
0.012% + 350 µV
20.0000
V
100 µV
>10
16
Ω
0.015% + 1.5 mV
200.000
V
1mV
>10
16
Ω
0.015% + 10 mV
TEMPERATURE COEFFICIENT (0
°–
18
°
C & 28
°–
40
°
C):
±(0.15
×
accuracy specification)/
°
C.
Current Measurement Accuracy (2- or 4-wire sense)
4
Accuracy (23
°
C ± 5
°
C)
Max.
Voltage
1 Year
Range
Resolution
*
Burden
5
±(%rdg + amps)
1.00000 pA
10 aA
< 1mV
1.0
% +
7 fA
10.0000 pA
100 aA
< 1mV
0.50 % +
7 fA
100.000 pA
1 fA
< 1mV
0.15 % + 30 fA
1.00000 nA
10 fA
< 1mV
0.050 % + 200 fA
10.0000 nA
100 fA
< 1mV
0.050 % +
2 pA
100.000 nA
1 pA
< 1mV
0.050 % + 20 pA
1.00000 µA
10 pA
< 1mV
0.050 % + 300 pA
10.0000 µA
100 pA
< 1mV
0.050 % +
2 nA
100.000 µA
1 nA
< 1mV
0.025 % +
6 nA
1.00000 mA
10 nA
< 1mV
0.027 % + 60 nA
10.0000 mA
100 nA
< 1mV
0.035 % + 600 nA
100.000 mA
1 µA
< 1mV
0.055 % +
6 µA
TEMPERATURE COEFFICIENT (0
°–
18
°
C & 28
°–
40
°
C):
±[(0.15
×
accuracy specification) + 1fA]/
°
C.
INPUT CURRENT:
<3fA at 23
°
C, <40% RH; typically ±0.5fA/
°
C around 23
°
C, <40% RH.
ADDITIONAL MEASURE SPECIFICATIONS
OUTPUT SETTLING TIME (typical to 10% of final value):
<2s, 1pA and 10pA ranges; <50ms, 100pA through 10nA ranges; <5ms, 100nA through
100mA ranges.
CURRENT NOISE:
When observed over 1 minute intervals, peak to peak noise will be within 400aA during 90% of the intervals using Autofilter
(5s 10% to 90% rise time), with triax connectors capped, Autozero OFF, Source Delay = 0, on the 1pA range for at least 3 minutes.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......