SCPI Command Reference
17-17
Table 17-7
STATus command summary
Command
Description
Default
parameter
SCPI
:STATus
:MEASurement
[:EVENt]?
:ENABle <NDN>
or <NRf>
:ENABle?
:CONDition?
:OPERation
[:EVENt]?
:ENABle <NDN>
or <NRf>
:ENABle?
:CONDition?
:QUEStionable
[:EVENt]?
:ENABle <NDN>
or <NRf>
:ENABle?
:CONDition?
:PRESet
:QUEue
[:NEXT]?
:ENABle <list>
:ENABle?
:DISable <list>
:DISable?
:CLEar
Control measurement event registers:
Read the event register.
6
Program the enable register.
Read the enable register.
6
Read the condition register.
6
Control operation status registers:
Read the event register.
6
Program the enable register.
Read the enable register.
6
Read the condition register.
6
Control questionable status registers:
Read the event register.
6
Program the enable register.
Read the enable register.
6
Read the condition register.
6
Return status registers to default states.
Path to access error queue:
Read the most recent error message.
Specify error and status messages for error queue.
Read the enabled messages.
Specify messages not to be placed in error queue.
Read the disabled messages.
Clears all messages from error queue.
Note 1
Note 2
Note 3
Note 2
Note 3
Note 2
Note 3
Note 4
Note 5
Note 5
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
Notes:
1. Commands in this subsystem are not affected by *RST and :SYSTem:PREset. The effects of cycling power, *CLS and
:STATus:PRESet, are explained by the following notes.
2. Event Registers
—
Power-up and *CLS clears all bits.:STATus:PRESet has no effect.
3. Enable Registers
—
Power-up and :STATus:PRESet clears all bits. *CLS has no effect. Accepts the SCPI 1995.0 mandated
(non-decimal numeric) format (#H, #Q, or #B).
4. Error Queue
—
Power-up and *CLS clears all bits of the registers.
5. Error Queue Messages
—
Power-up clears list of messages. *CLS and :STATus:PRESet have no effect.
6. Register Query Commands
—
The format for the response messages (ASCII, hexadecimal, octal, or binary) depends on which
data format is presently selected. See the :FORMat:SREGister command.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......