17-68
SCPI Command Reference
If a manual source range is presently selected, then the speci
fi
ed amplitude
cannot exceed that range. For example, if the V-Source is on the 2V range
(auto range disabled), you will not be able to set the V-Source amplitude to
3V. In auto range, the amplitude can be set to any level that is within the
capabilities of the source.
The MINimum and MAXimum parameters are only valid if the highest
source range is presently selected. Sending the MINimum or MAXimum
parameters on a lower source range will generate error -221 (Setting
Con
fl
ict).
:TRIGgered[:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel]:TRIGgered[:AMPLitude] <n>
Set
fi
xed I-Source amplitude when trig-
gered
:SOURce[1]:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <n>
Set
fi
xed V-Source amplitude when
triggered
Parameters
<n> =
-105e-3 to 105e-3
Set I-Source amplitude (amps)
-210 to 210
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-105e-3A or -210V
MAXimum
+105e-3A or +210V
Query
:TRIGgered?
Query triggered amplitude for
fi
xed source
:TRIGgered? DEFault
Query *RST default amplitude
:TRIGgered? MINimum
Query lowest allowable amplitude
:TRIGgered? MAXimum Query highest allowable amplitude
Description
This command performs the same as the [:IMMediate][:AMPLitude] com-
mand except that the amplitude is not updated immediately.
With this command, the amplitude is updated when the SourceMeter is
triggered to perform a source-measure operation. For example, if the
instrument is waiting in the trigger layer for an external trigger, the
amplitude of the source will not update until that external trigger is received
by the SourceMeter. See Section 10,
Trigger model
for details on trigger
model operation.
The MINimum and MAXimum parameters are only valid if the highest
source range is presently selected. Sending the MINimum or MAXimum
parameters on a lower source range will generate error -221 (Setting
Con
fl
ict).
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......