18-28
Performance Verification
4.
Verify ohms measurement accuracy for each of the resistance values listed in
Table 18-14. For each measurement:
•
Set the resistance calibrator output to the nominal resistance or closest available
value.
NOTE
It may not be possible to set the resistance calibrator to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
•
Select the appropriate ohms measurement range with the RANGE keys.
•
Verify that the Model 6430 resistance reading is within the limits given in the
table.
2G
Ω
-200G
Ω
range accuracy
1.
With the power off, connect the Remote PreAmp IN/OUT HIGH jack directly to the
Model 5156 OUTPUT jack as shown in Figure 18-9. (Use male triax-to-triax adapters.
Also, be sure to remove the link between Model 5156 SHIELD and CHASSIS, and
connect the BNC shorting cap to the resistance cap being used.
)
Table 18-14
Remote PreAmp 20
Ω
-200M
Ω
range measurement accuracy limits
Model 6430 range
Calibrator resistance
1
Model 6430 resistance reading limits
2
(1 year, 18°-28°C)
20
Ω
200
Ω
2k
Ω
20k
Ω
200k
Ω
2M
Ω
20M
Ω
200M
Ω
19
Ω
190
Ω
1.9k
Ω
19k
Ω
190k
Ω
1.9M
Ω
19M
Ω
100M
Ω
18.9784 to 19.0216
Ω
189.824 to 190.176
Ω
1.89845 to 1.90155k
Ω
18.9850 to 19.0150k
Ω
189.814 to 190.186k
Ω
1.89814 to 1.90186M
Ω
18.9829 to 19.0171M
Ω
99.905
to 100.095M
Ω
1
Nominal resistance value.
2
Reading limits based on Model 6430 normal accuracy speci
fi
cations and nominal resistance values. If actual
resistance values differ from nominal values shown, recalculate reading limits using actual calibrator resis-
tance values and Model 6430 normal accuracy speci
fi
cations. See
Verification limits
earlier in this section for
details.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......