Performance Verification
18-13
1.
With the power off, connect the digital multimeter to the Model 6430 INPUT/OUTPUT
jacks as shown in Figure 18-3. (Connect Model 6430 INPUT/OUTPUT HI to DMM
AMPS input; INPUT/OUTPUT LO to INPUT LO.)
2.
Select the multimeter DC current measuring function.
3.
Press the Model 6430 SOURCE I key to source current, and make sure the source out-
put is turned on.
4.
Verify output current accuracy for the currents listed in Table 18-5. For each test point:
•
Select the correct source range.
•
Set the Model 6430 output current to the correct value.
•
Verify that the multimeter reading is within the limits given in the table.
5.
Repeat the procedure for negative output currents with the same magnitudes as those
listed in Table 18-5.
Table 18-5
Mainframe output current accuracy limits
Model 6430
source range
Model 6430
output current setting
Output current limits
(1 year, 18°-28°C)
1µA
10µA
100µA
1mA
10mA
100mA
1.00000µA
10.0000µA
100.000µA
1.00000mA
10.0000mA
100.000mA
0.99905 to 1.00095µA
9.9947 to 10.0053µA
99.949 to 100.051µA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
Model 6430
Digital Multimeter
Input LO
Amps
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V,
Ω
,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Figure 18-3
Connections for
mainframe current
verification tests
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......