17-32
SCPI Command Reference
:ACQuire
:CALCulate2:NULL:ACQuire
Automatically acquire REL value
Description
This command automatically acquires the null offset value. The next avail-
able reading will become the offset value.
:STATe <b>
:CALCulate2:NULL:STATe <b>
Control null offset
Parameters
<b> =
1 or ON
Enable null offset
0 or OFF
Disable null offset
Query
:STATe?
Query state of null offset
Description
This command is used to enable or disable null offset. When enabled, the
CALC2 reading will include the null offset value. See :OFFSet. When dis-
abled, CALC2 will not include the null offset.
Read CALC2
:DATA?
:CALCulate2:DATA?
Read CALC2
Description
This command is used to acquire all the readings used for the CALC2 limit
tests. Note that if null offset is enabled, then the CALC2 readings will
include the null offset value. See
Null Feed Reading
.
At least one of the limit tests have to be enabled to a acquire limit test read-
ings. See
Configure and control limit tests
; :STATe.
NOTE
See Appendix C for a detailed explanation on how data flows through the various
operation blocks of the SourceMeter. It clarifies the type of readings that are
acquired by the various commands to read data.
:LATest?
:CALCulate2:DATA:LATest?
Read latest CALC2 data
Description
This command works exactly like CALC2:DATA, except it returns only the
latest null offset or limit result.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......