SCPI Command Reference
17-33
Con
fi
gure and control limit tests
:COMPliance:FAIL <name>
:CALCulate2:LIMit[1]:COMPliance:FAIL <name>
Parameters
<name> =
IN
Fail Limit 1 test when unit goes into
compliance
OUT
Fail Limit 1 test when unit comes out of
compliance
Query
:FAIL?
Query when Limit 1 test failure occurs
Description
This command is used to specify the condition that will cause Limit 1 test to
fail. With IN speci
fi
ed, the test will fail when the SourceMeter goes into
compliance. With OUT speci
fi
ed, the test will fail when the SourceMeter
comes out of compliance.
[:DATA] <n>
:CALCulate2:LIMitx:LOWer[:DATA] <n>
Specify lower LIMIT x (x = 2, 3, 5-12)
:CALCulate2:LIMitx:UPPer[:DATA] <n>
Specify upper LIMIT x (x = 2, 3, 5-12)
Parameters
<n> =
-9.999999e20 to
Specify limit value
9.999999e20
DEFault
Set speci
fi
ed lower limit to -1
Set speci
fi
ed upper limit to 1
MINimum
Set speci
fi
ed limit to -9.999999e20
MAXimum
Set speci
fi
ed limit to +9.999999e20
Query
:UPPer?
Query speci
fi
ed upper limit
:UPPer? DEFault
Query *RST default upper limit
:UPPer? MINimum
Query lowest allowable upper limit
:UPPer? MAXimum
Query largest allowable upper limit
:LOWer?
Query speci
fi
ed lower limit
:LOWer? DEFault
Query *RST default lower limit
:LOWer? MINimum
Query lowest allowable lower limit
:LOWer? MAXimum
Query largest allowable lower limit
Description
These commands are used to set the upper and lower limits for LIMIT 2,
LIMIT 3, and LIMIT 5 through LIMIT 12 tests. The actual limit depends on
which measurement function is currently selected. For example, a limit
value of 1µ is 1µA for the amps function and 1µV for the v olts function. A
limit value is not range sensitive. A limit of 2 for volts is 2V on all measure-
ment ranges.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......