11-10
Limit Testing
Binning systems
The SourceMeter can be used with a component handler to perform binning operations on
DUT packages. With this system, you can test single-element devices (i.e., resistor). Adding a
scanner to the system allows binning operations on multiple-element DUT packages. See
Limit
test programming example
at the end of this section.
Handler interface
The SourceMeter is interfaced to a handler via the Digital I/O port as shown in Figure 11-6.
The I/O port has four lines for output signals and one line for input signals. The output lines are
used to send the test pass/fail signal(s) to the handler to perform the binning operation.
Digital I/O connector
These digital I/O lines are available at the DB-9 Digital I/O connector on the rear panel of
the SourceMeter. A custom cable using a standard female DB-9 connector is required for con-
nection to the SourceMeter. See
Digital I/O port
in Section 12 for more information.
Digital output lines
The four output lines output a speci
fi
c bit pattern based on the pass/fail results of the various
limit tests. (See
Types of limits
earlier in this section). In the 3-bit output mode, Line 4 can also
be used either as an EOT (End of Test) or BUSY signal depending on the END OF TEST
mode. (See
Configuring limit tests
later in this section.)
SourceMeter
Handler
Out 1
Out 2
Out 3
Out 4
Gnd
Dig I/O
1
6
5
9
+5V
Input (SOT)
Line 1
Line 2
Line 3
Line 4 (EOT or BUSY)
SOT Strobe Line
/INT
Gnd
Figure 11-6
Handler interface
connections
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......