Digital I/O Port, Interlock, and Output Configuration
12-3
EOT/BUSY line
Line 4 can be used for a normal bit pattern, EOT (End-of-Test), or BUSY signal, depending
on the selected END OF TEST mode.
NOTE
See Section 11 for details on performing limit tests and Section 10 for information
on programming the SourceMeter to respond to the SOT (start-of-test) pulse from a
handler.
+5V output
The Digital I/O Port provides a +5V output that can be used to drive external logic circuitry.
Maximum current output for this line is 300mA. This line is protected by a self-resetting fuse
(one hour recovery time).
Digital output con
fi
guration
There are two basic methods to connect external components to the digital output lines, sink
operation and source operation.
Sink operation
Figure 12-2 shows the basic output con
fi
guration for sink operation. Note that the external
relay coil is connected between the digital output line (pins 1 to 4) and +5V (pin 7). With this
con
fi
guration, the digital output line must be set LO to energize the relay, and the maximum
sink current is 500mA.
SourceMeter
External
Relay
To other
Circuits
+5V
Maximum sink current: 500mA
Pin 7
Pins 1-4
Pin 9
Digital I/O
Port
Figure 12-2
Sink operation
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......