![background image](http://html.mh-extra.com/html/keithley/6430/6430_instruction-manual_661765165.webp)
Relative and Math
7-9
Table 7-4
Voltage coefficient programming example
Command
Description
*RST
:SENS:FUNC:ON:ALL
:SENS:RES:MODE MAN
:SOUR:FUNC VOLT
:SOUR:VOLT:STAR 10
:SOUR:VOLT:STOP 50
:SOUR:VOLT:MODE SWE
:SOUR:SWE:POIN 2
:TRIG:COUN 2
:CALC:MATH:NAME “VOLTCOEF”
:CALC:STAT ON
:OUTP ON
:INIT
:CALC:DATA?
Reset unit to GPIB defaults.
Enable all sense functions.
Manual resistance mode.
Volts source function.
10V start voltage.
50V stop voltage.
Volts sweep mode.
Sweep points = 2.
Trigger count = 2.
Select voltage coefficient math expression.
Enable math.
Turn on output.
Trigger sweep.
Request voltage coefficient data.
Triax Cable
HI
LO
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
Connect to REMOTE
PreAmp connector on rear
panel of mainframe
Resistor
Under
Test
Preamp Cable
IN/OUT
HIGH
Figure 7-2
Connections for
voltage coefficient
tests
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......