DM-Series Field Calibration Guide 

 

 

There are 3 keys, M, S, and Z.   

  Key M is to save data and settings. 

  Key S for data setting and numerical increment. 

  Key Z for digit shifting and option. 

 

Setting data using M, S, and Z keys

The following is a step by step walkthrough on navigating the DM-Series using the 3 
programming keys. Please note: Some calibrations require programming not available through 
basic button functions. For calibrations that require programming please contact Eagle Eye 
Power Solutions at our toll-free line: 1-877-805-3377 or email us at 

[email protected]

  

Interface 1 

– Code input screen 

Press Z for up to 3 seconds and you enter into data setting interface code 1. Press Z key to shift 
the flash digit to the last space. 

Press S key to change the digit to 2, then press Z key. There is one flash arrow on the lower 
left, press the S key. Turn to the unit setting interface of operation code 2. 

Interface 2 

– Setting units 

The current unit is now shown, press S key and the unit option begins flashing, press S key 
again and the unit changes. Continuing to press S key subsequent times will cycle through 
available units. Choose the unit you would like to use then press M key to save. 

Interface 3 

– Setting lower limit range 

After you save the measuring unit you will enter into the interface operation code 3.  If you do 
not need to set the lower limit press M key to save the setting. 

Interface 4 

– Setting upper limit range 

After you save you will then enter into interface operation code 4. The current upper limit range 
should be either 1, 2, or 3.  Press the S key to set the upper range. Once complete press M key 
to save. 

 

 