Sweep Operation
9-17
Step 4: Turn output on.
Press the ON/OFF OUTPUT key to turn the output on (OUTPUT indicator turns on). The
SourceMeter will output the programmed bias level.
Step 5: Run sweep.
To run the sweep, press the SWEEP key. After the sweep is completed, turn the output off by
pressing the ON/OFF OUTPUT key.
Step 6: Read buffer.
Use the RECALL key to access the source-measure readings stored in the buffer. Use the
TOGGLE to display statistical information.
Performing a source memory sweep
Step 1: Store setups in source memory.
Store instrument setups in source memory as follows:
1.
Configure the SourceMeter for various desired operating modes such as source, mea-
sure, delay, and/or math expression operation. See Table 9-2 for settings that can be
stored in each source memory location.
2.
Press MENU to display the MAIN MENU:
•
Select SAVESETUP.
•
Select SOURCE MEMORY.
•
Select SAVE.
•
Use the
▲
and
▼
keys, and the cursor keys to display the desired memory location,
and press ENTER.
•
Use the EXIT key to back out of the menu structure.
3.
Repeat Steps 1 and 2 for all points in the sweep.
Step 2: Configure sweep
Configure the sweep as follows:
1.
Press CONFIG then SWEEP.
2.
Select TYPE, then press ENTER.
3.
Select SRC MEMORY, then press ENTER.
4.
Use the menu selections to enter the desired START memory location and # POINTS
for the source memory sweep.
5.
From the CONFIGURE SWEEPS menu, select SWEEP COUNT, press ENTER, then
choose FINITE or INFINITE as desired.
6.
Again from the CONFIGURE SWEEPS menu, choose SOURCE RANGING, press
ENTER, then select BEST FIXED, AUTO RANGE, or FIXED as appropriate.
7.
Press EXIT to return to normal display.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......