Getting Started
1-13
Power-up sequence
On power-up, the SourceMeter performs self-tests on its EPROM and RAM and momen-
tarily lights all segments and annunciators. If a failure is detected, the instrument momentarily
displays an error message, and the ERR annunciator turns on. Error messages are listed in
Appendix B.
NOTE
If a problem develops while the instrument is under warranty, return it to Keithley
Instruments, Inc., for repair.
If the instrument passes the self-tests, the firmware revision levels are displayed. For
example:
REV A01 A02
where: A01 is the main board ROM revision.
A02 is the display board ROM revision.
Also displayed is the line frequency (50, 60, or 400Hz). (If the wrong frequency is dis-
played, it can be manually set as covered below). The communication interface status is briefly
displayed. If the IEEE-488 bus is the presently selected interface, the identification message
will include the primary address. For example, if the primary address is 24 (factory default), the
“IEEE Addr=24” message is displayed. If the RS-232 interface is selected, the “RS-232” mes-
sage is displayed.
After the power-up sequence, the instrument goes to its normal display state with the output
off (red OUTPUT indicator light off). With the output off, the “OFF” message is displayed, and
dashes replace the reading. The “OFF” message indicates that the NORMAL output-off state is
selected. See Section 12,
Output configuration
for details.
System identi
fi
cation
To obtain the serial number and revision information, use the MENU/GENERAL/SERIAL #
selection or the *IDN? query via remote.
Line frequency setting
At the factory, the SourceMeter is configured to sense the power line frequency and auto-
matically select the frequency setting. If, however, the line power source is noisy, the Source-
Meter may select the wrong setting on power-up. If this situation occurs, noisy measurement
readings will result, and accuracy may be affected. You can manually set the line frequency
from the front panel or via remote as covered below.
Front panel line frequency
Set the line frequency from the front panel as follows:
1.
Press the MENU key to display MAIN MENU.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......