SCPI Command Reference
17-53
:TRIPped?
:OUTPut[1]:INTerlock:TRIPped?
Description
This query command is used to determine if the enabled interlock has been
tripped. The tripped condition (
“
1
”
) means that the source can be turned on
(interlock line at logic low level).
A
“
0
”
will be returned if the source cannot be turned on (interlock line at
logic high level).
Output-off states
:SMODe
:OUTPut[1]:SMODe <name>
Select output-off mode
Parameters
<name> =
NORMal
Normal output-off state
ZERO
Zero output-off state
GUARd
Guard output-off state
Query
:SMODe?
Query output off mode
Description
This command is used to select the output-off state of the SourceMeter.
With NORMal selected (which is the default), the V-Source is selected and
set to 0V when the output is turned off. Compliance is set to 0.5% full scale
of the present current range.
In the ZERO output-off state when the V-Source OUTPUT is turned off, the
V-Source is set to 0V and current compliance is not changed. When the
I-Source OUTPUT is turned off, the V-Source mode is selected and set to
0V. Current compliance is set to the programmed Source I value or to 0.5%
full scale of the present current range, whichever is greater.
The ZERO output-off state is typically used with the V-Source and Output
Auto-On (see the :SOURce1:CLEar:AUTO command) to generate voltage
waveforms that alternate between 0V and the programmed output-on
voltage.
With GUARd selected, the I-Source is selected and set to 0A. Voltage com-
pliance is set to 0.5% full scale of the present voltage range. This output-off
state should be used when performing 6-wire guarded ohms measurements
or for any other load that uses an active source.
NOTE
For more information on output-off states, see Section 12, “Output configuration.”
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......