13-6
Remote Operations
3.
Connect any additional connectors from other instruments as required for your
application.
4.
Make sure the other end of the cable is properly connected to the controller. Most con-
trollers are equipped with an IEEE-488 style connector, but a few may require a differ-
ent type of connecting cable. See your controller's instruction manual for information
about properly connecting to the IEEE-488 bus.
NOTE
You can only have 15 devices connected to a IEEE-488 bus, including the controller.
The maximum cable length is either 20 meters or two meters multiplied by the num-
ber of devices, whichever is less. Not observing these limits may cause erratic bus
operation.
Primary address
The SourceMeter ships from the factory with a GPIB primary address of 24. When the unit
powers up, it momentarily displays the primary address. You can set the address to a value from
0 to 30, but do not assign the same address to another device or to a controller that is on the
same GPIB bus (controller addresses are usually 0 or 21).
The primary address can be checked and/or changed from the COMMUNICATIONS/GPIB
option of the Main Menu. See Section 1,
Main menu
. This menu option also allows you to
select the 488.1 or SCPI protocol (see Appendix G).
General bus commands
General commands are those commands, such as DCL, that have the same general meaning
regardless of the instrument. Table 13-1 lists the general bus commands.
Table 13-1
General bus commands
Command
Effect on SourceMeter
REN
IFC
LLO
GTL
DCL
SDC
GET
SPE, SPD
Goes into remote when next addressed to listen.
Goes into talker and listener idle states.
LOCAL key locked out.
Cancel remote; restore SourceMeter front panel operation.
Returns all devices to known conditions.
Returns SourceMeter to known conditions.
Initiates a trigger.
Serial polls the SourceMeter.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......