SCPI Command Reference
17-67
:AUTO <b>
:SOURce[1]:CURRent:RANGe:AUTO <b>
Select auto range for I-Source
:SOURce[1]:VOLTage:RANGe:AUTO <b>
Select auto range for V-Source
Parameters
<b> =
0 or OFF
Disable auto range
1 or ON
Enable auto range
Query
AUTO?
Query state of auto range
Description
This command is used to enable or disable auto range for the speci
fi
ed
source. When enabled, the instrument will automatically select the most
sensitive range for the speci
fi
ed source level. When disabled, the instrument
will use the range that the instrument is currently on.
Auto range will be disabled if a
fi
xed range is selected. See the previous
command.
Both *RST and :SYSTem:PREset enables source auto range. When the
SourceMeter goes into the local state, source auto range disables.
Set amplitude for
fi
xed source
[:IMMediate][:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude] <n> Set
fi
xed I-Source amplitude immediately
:SOURce[1]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <n> Set
fi
xed V-Source amplitude immediately
Parameters
<n> =
-105e-3 to 105e-3
Set I-Source amplitude (amps)
-210 to 210
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-105e-3A or -210V
MAXimum
+105e-3A or +210V
Query
:CURRent?
Query programmed amplitude of I-Source
:CURRent? DEFault
Query *RST default amplitude
:CURRent? MINimum
Query lowest allowable amplitude
:CURRent? MAXimum
Query highest allowable amplitude
:VOLTage?
Query programmed amplitude of V-Source
:VOLTage? DEFault
Query *RST default amplitude
:VOLTage? MINimum
Query lowest allowable amplitude
:VOLTage? MAXimum
Query highest allowable amplitude
Description
This command is used to immediately update the amplitude of a
fi
xed
source. This command is not valid if the list or sweep mode is selected.
NOTE
The sourcing :MODE command is used to select a fixed source. See “Select sourcing
mode.”
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......