SCPI Command Reference
17-69
Set voltage limit
[:LEVel] <n>
:SOURce[1]:VOLTage:PROTection[:LEVel] <n>
Set voltage limit for V-Source
Parameters
<n> =
-210 to 210
Specify V-Source limit
20
Set limit to 20V
40
Set limit to 40V
60
Set limit to 60V
80
Set limit to 80V
100
Set limit to 100V
120
Set limit to 120V
160
Set limit to 160V
161 to 210
Set limit to NONE
NONE
Set limit to 210V
DEFault
Set limit to 210V (NONE)
MINimum
Set limit to 20V
MAXimum
Set limit to 210V (NONE)
Query
[:LIMit]?
Query limit level
[:LIMit]? DEFault
Query *RST default limit
[:LIMit]? MINimum
Query lowest allowable limit
[:LIMit]? MAXimum
Query highest allowable limit
Description
This command is used to set the Over Voltage Protection (OVP) limit for the
V-Source. The V-Source output will not exceed the selected limit. An
exception to this is a parameter value that exceeds 160V. Exceeding this
value allows the V-Source to output its maximum voltage. The OVP limit is
also enforced when in the I-Source Mode.
The limit parameter values are magnitudes and are in effect for both positive
and negative output voltage. You can express the limit as a positive or nega-
tive value.
If you specify a value that is less than the lowest limit, the lowest limit will
be selected. If you specify a value that is between limits, the lower limit will
be selected. For example, if you specify a value of 110, the 100V limit will
be selected.
WARNING
Even with the voltage protection limit set to the lowest value (20V),
NEVER touch anything connected to the terminals of the SourceMeter
when the OUTPUT is ON. Always assume that a hazardous voltage
(>30V rms) is present when the OUTPUT is ON.
To prevent damage to DUT (devices under test) or external circuitry, DO
NOT program the V-Source to levels that exceed the voltage protection
limit.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......