SCPI Command Reference
17-39
:MODE <name>
:CALCulate2:CLIMits:MODE <name>
Control Digital I/O port pass/fail output
Parameters
<name> =
GRADing
Output graded
“
pass/fail
”
pattern
SORTing
Output sorted
“
pass/fail
”
pattern
Query
:MODE?
Query Digital I/O pass/fail mode
Description
This command controls how limit calculations drive the Digital I/O lines. In
GRADing mode, a reading passes if it is within all of the hi/low limit toler-
ances enabled, assuming that it has passed LIMIT 1 compliance test
fi
rst.
The Digital I/O lines will be driven with the
fi
rst pattern of the
fi
rst compli-
ance, hi, or low failure. Otherwise, the CALC2:CLIM:PASS:SOUR2 pat-
tern will be output.
In SORTing mode, a reading will fail if it fails the compliance test, or is not
within any of the Digital I/O Bands. If the tests pass and only LIMIT 1 is
enabled, the CALC2:CLIM:PASS:SOUR2 pattern will be output. Other-
wise, the
fi
rst limit test band that passes will output its LOW:SOUR2 pattern
(UPP:SOUR2 patterns will be ignored). If LIMIT1 fails, its SOUR2 pattern
will be output. If no LIMIT2, 3, 5-12 limit passes, the
CALC2:CLIM:FAIL:SOUR2 pattern will be output.
Clear test results
[:IMMediate]
:CALCulate2:CLIMits:CLEar[:IMMediate]
Clears test results and resets Digital I/O
Port
Description
This command clears the test results (pass or fail) of the limit tests and
resets the output lines of the Digital I/O port back to the :SOURce2:TTL
settings. See
SOURce2
.
:AUTO <b>
:CALCulate2:CLIMits:CLEar:AUTO <b>
Control auto-clear for test results
Parameters
<b> =
1 or ON
Enable auto-clear
0 or OFF
Disable auto-clear
Query
:AUTO?
Query state of auto-clear
Description
With auto-clear enabled, test results will clear and the output lines of the
Digital I/O port will reset when the :INITiate command is sent to start a new
test sequence.
When disabled, you must use :IMMediate to perform the clear actions.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......