17-106 SCPI Command Reference
Description
When TLINk is the selected Trigger Layer control source, and an event
detector in the Trigger Layer is enabled, operation will hold up at that detec-
tor until an input trigger is received via the Trigger Link. When the event
detector is disabled, operation will not hold up. It continues on and performs
the appropriate action.
A Trigger Layer event detector is enabled by including the parameter name
in the event list for the INPut command. For example, to enable the Source
Event Detector and Measure Event Detector, send the following command:
:TRIGger:INPut SOURce, SENSe
The Delay Event Detector will be disabled since the DELay parameter is not
included in the above event list.
NOTE
To disable all the Trigger Layer event detectors, the NONE parameter must be sent
alone (i.e. trigger:input none). If it is listed with any of the other parameters, NONE
will be ignored.
:ILINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCON
fi
gure]:ILINe <NRf>
Select input line; arm layer
:TRIGger[:SEQuence[1]][:TCON
fi
gure]:ILINe <NRf>
Select input line; trigger layer
Parameters
<NRf> =
1
Line #1
2
Line #2
3
Line #3
4
Line #4
Query
:ILINe?
Query input trigger line
Description
This command is used to select input lines for the Trigger Link. For normal
operation, Trigger Link input and output (see :OLINe) should not share the
same line.
:OLINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCON
fi
gure]:OLINe <NRf> Select output line; arm layer
:TRIGger[:SEQuence[1]][:TCON
fi
gure]:OLINe <NRf>
Select output line; trigger layer
Parameters
<NRf> =
1
Line #1
2
Line #2
3
Line #3
4
Line #4
Query
:OLINe?
Query output trigger line
Description
This command is used to select output lines for the Trigger Link. For normal
operation, Trigger Link input and output (see :ILINe) should not share the
same line.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......