SCPI Command Reference 17-107
OUTPut <event list>
:ARM[:SEQuence[1]][LAYer[1]][:TCON
fi
gure]:OUTPut <event list>
Arm layer events
:TRIGger[:SEQuence[1]][:TCON
fi
gure]:OUTPut <event list>
Trigger layer events
Parameters
Arm Layer Triggers
<event list >: TENTer
Trigger on entering trigger layer
TEXit
Trigger on exiting trigger layer
NONE
Disable arm layer output trigger
Trigger Layer Triggers
<event list>: SOURce
Output trigger after source level is set
DELay
Output trigger after delay period
SENSe
Output Trigger after measurement
NONE
Disable trigger layer triggers
NOTE
Each event in the list must be separated by a comma (i.e., :arm:output source, delay,
sense).
Query
:OUTPut?
Query output trigger event(s)
Description
This command is used to specify when trigger pulses occur on the speci
fi
ed
output trigger line of the Trigger Link (:OLINe).
Arm Layer Triggers
—
With TEXit selected, an output trigger will occur
when exiting the trigger layer. With TENTer selected, an output trigger will
occur when entering the trigger layer. With NONE selected, the arm layer
output trigger is disabled.
Trigger Layer Triggers
—
You can specify from one to all three events. Each
event in the list must be separated by a comma (,).
The SOURce, DELay and MEASure events refer to the Source-Delay-
Measure (SDM) cycle. This is the Device Action in the Trigger Model. See
Figures 10-7 and 10-8. With SOURce speci
fi
ed, an output trigger occurs
after the source is set. With DELay speci
fi
ed, an output trigger occurs after
the delay period. With MEASure speci
fi
ed, an output trigger occurs after the
measurement.
NOTE
When disabling triggers, the NONE parameter must be sent alone (i.e., trig:outp
none). If it is listed with any of the other event parameters, NONE will be ignored.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......