1-14
Getting Started
2.
Using the right arrow key, select AD-CTRL then press ENTER to display A/D
CONTROLS.
3.
Select LINE-FREQ, then press ENTER to display LINE FREQUENCY.
4.
Place the cursor on 50Hz, 60Hz, or AUTO, and press ENTER. (Use 50Hz for 400Hz
operation.)
5.
Press EXIT to return to normal display.
Remote command line frequency
Table 1-1 summarizes remote commands used to control line frequency. See Section 17 for
additional information on using these commands.
Programming examples
50 or 400Hz operation:
:SYST:LFR 50
Auto frequency operation:
:SYST:LFR:AUTO ON
Fuse replacement
A rear panel fuse protects the power line input of the SourceMeter. If the line fuse needs to
be replaced, perform the following steps:
1.
The fuse is located in a drawer above the AC receptacle. See Figure 1-2. At the bottom
of the fuse drawer is a small tab. At this location, use a small bladed screwdriver to pry
the fuse drawer open.
2.
Slide the fuse drawer out to gain access to the fuse. Note that the fuse drawer does not
pull all the way out of the power module.
CAUTION
For continued protection against fire or instrument damage, replace the
fuse only with the type and rating listed. If the instrument repeatedly
blows fuses, locate and correct the cause of the problem before replacing
the fuse.
3.
Snap the fuse out of the drawer and replace it with the same type (250V, 2.5A,
5
×
20mm). The Keithley part number is FU-72.
4.
Push the fuse drawer back into the power module.
Table 1-1
Line frequency remote commands
Commands
Description
:SYSTem:LFRequency <freq>
:SYSTem:LFRequency:AUTO <state>
Select line frequency (freq = 50 or 60).
Enable/disable auto frequency (state = ON or OFF).
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......