6-16
Range, Digits, Speed, and Filters
Filter control
When filtering is being applied to the input signal, the FILT annunciator will be on. When
Auto Filter is enabled, the FILT annunciator will turn on to indicate that the Auto Filter config-
uration is being applied.
The FILTER key is used to control filtering. Pressing FILTER turns on the FILT annunciator
to indicate that the filter configuration is being applied to the input. Pressing FILTER a second
time turns the FILT annunciator off to indicate that filtering is turned off.
Remote
fi
lter programming
Filter commands
Table 6-8 summarizes filter commands. See Section 17 for more details.
NOTE
When Auto Filter is enabled, the filter configuration cannot be changed. Therefore,
the commands in Table 6-8 for the Repeat Filter, Median Filter and Moving Filter
are no-operations with Auto Filter on.
Table 6-8
Filter commands
Commands
Description
for Auto Filter:
[:SENSe]:AVERage:AUTO <state>
for Repeat Filter:
[:SENSe]:AVERage:REPeat:COUNt <n>
[:SENSe]:AVERage:REPeat[:STATe] <state>
for Median Filter:
[:SENSe]:MEDian:RANK <NRf>
[:SENSe]:MEDian[:STATe] <state>
for Moving Filter:
[:SENSe]:AVERage:COUNt <n>
[:SENSe]:AVERage[:STATe] <state>
[:SENSe]:AVERage:ADVanced:NTOLerance <NRf>
[:SENSe]:AVERage:ADVanced[:STATe] <state>
Enable/disable Auto Filter (state = ON or OFF).
Set Repeat Filter count (n = count, 1 to 100).
Enable/disable Repeat Filter (state = ON or OFF).
Set Median Filter rank (NRf = rank, 0 to 5).
Enable/disable Median Filter (state = ON or OFF).
Set Moving Filter count (n = count, 1 to 100).
Enable/disable Moving Filter (state = ON or OFF).
Set Advanced Filter noise window in % (NRf = noise
window, 0 to 105).
Enable/disable Advanced Filter (state = ON or OFF).
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......