Ohms Measurements
4-5
3.
Select measurement range.
Use the RANGE
▲
and
▼
keys to select a range appropriate for the expected ohms
reading, or use autorange by pressing AUTO. When using manual ranging, selecting the
most sensitive (lowest) range provides the best accuracy. Autorange automatically goes
to the most sensitive range.
4.
Turn output on.
Turn the output on by pressing the ON/OFF OUTPUT key. The OUTPUT indicator will
turn on to indicate the output is on.
5.
Observe reading on display.
The SourceMeter will go into compliance if you exceed the maximum ohms measure-
ment range.
6.
Turn output off.
When finished, turn the output off by pressing the ON/OFF OUTPUT key. The
OUTPUT indicator light will turn off.
Manual ohms measurements
Perform the following steps to perform manual ohms measurements.
NOTE
The following procedure assumes that the SourceMeter is already connected to the
DUT as explained in Section 2.
WARNING
To prevent electric shock, do not make or break connections to the Source-
Meter with the output on. If on, press the ON/OFF OUTPUT key to turn
the output off.
1.
Select ohms measurement function.
Press MEAS
Ω
to select the ohms measurement function.
2.
Select manual ohms measurement method.
•
Press CONFIG then
Ω
.
•
Select SOURCE, then press ENTER.
•
Select MANUAL, then press ENTER.
•
Press EXIT to return to normal display.
3.
Configure source.
For manual ohms, you can Source I or Source V at the user-programmed output level.
The lowest allowable compliance limit is based on the load and the source value. For
example, if sourcing 1V to a 1k
Ω
resistor, the lowest allowable current compliance is
1mA (1V/1k
Ω
= 1mA). Setting a limit lower than 1mA will place the source in compli-
ance. Refer to steps 1 and 2 of Section 3,
Basic source-measure procedure
to configure
the source.
NOTE
Use the V-Source for manual ohms measurements when high-speed settling is
required (i.e., production testing).
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......