19-8
Calibration
8.
Note and record the DMM reading, and then adjust the Model 6430 display to agree
exactly with the actual DMM reading. Use the up and down arrow keys to select the
digit value, and use the left and right arrow keys to choose the digit position (or use the
number keys, 0-9, +/-). Note that the display adjustment range is within ±10% of the
present range.
9.
After adjusting the display to agree with the DMM reading, press ENTER. The instru-
ment will then display the following:
V-CAL
Press ENTER to 000.00mV
10.
Press ENTER. The Model 6430 will source 0mV and at the same time display the
following:
DMM RDG: +000.0000mV
Use
,
▲
,
▼
,
, ENTER or EXIT.
11.
Note and record the DMM reading, and then adjust the Model 6430 display to agree
with the actual DMM reading. Note that the display value adjustment limits are within
±1% of the present range.
12.
After adjusting the display value to agree with the DMM reading, press ENTER. The
unit will then display the following:
V-CAL
Press ENTER to Output -200.00mV
13.
Press ENTER. The Model 6430 will source -200mV and display the following:
DMM RDG: -200.0000mV
Use
,
▲
,
▼
,
, ENTER or EXIT.
14.
Note and record the DMM reading, and then adjust the Model 6430 display to agree
with the DMM reading. Again, the maximum display adjustment is within ±10% of the
present range.
15.
After adjusting the display value to agree with the DMM reading, press ENTER and
note that the instrument displays:
V-CAL
Press ENTER to 000.00mV
16.
Press ENTER The Model 6430 will source 0mV and simultaneously display the
following:
DMM RDG: +000.0000mV
Use
,
▲
,
▼
,
, ENTER or EXIT.
17.
Note and record the DMM reading, and then adjust the display to agree with the DMM
reading. Once again, the maximum adjustment is within ±1% of the present range.
18.
After adjusting the display to agree with the DMM reading, press ENTER to complete
calibration of the present range.
19.
Press EXIT to return to normal display, and then select the 2V source range. Repeat
steps 2 through 18 for the 2V range.
20.
After calibrating the 2V range, repeat the entire procedure for the 20V and 200V ranges
using Table 19-3 as a guide. Be sure to select the appropriate source range with the
EDIT and RANGE keys before calibrating each range.
21.
Press EXIT as necessary to return to normal display.
▲
▲
▲
▲
▲
▲
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......