SCPI Command Reference
17-31
CALCulate2
Con
fi
gure and control limit tests
The following commands are used to con
fi
gure and control the three limit tests for DUT.
When used with a handler to provide binning operations, communication between the
SourceMeter and the handler is provided via the Digital I/O port. Many control aspects of the
digital output lines are performed from the SOURce2 Subsystem. These control aspects include
setting and clearing the digital output lines, and setting pulse width. See
SOURce2
for details.
Select input path
:FEED <name>
:CALCulate2:FEED <name>
Select input path for limit tests
Parameters
<name> = CALCulate[1]
Use result of CALC1
VOLTage
Use measured voltage reading
CURRent
Use measured current reading
RESistance
Use measured resistance reading
Query
:FEED?
Query input path for limit tests
Description
This command is used to select the input path for the limit tests. With
CALCulate[1] selected, the speci
fi
ed limits will be compared to the result of
CALC1. With VOLTage selected, limits will be compared to the voltage
measurement. With CURRent or RESistance selected, limits will be
compared with the respective current or resistance measurement.
Null feed reading
:OFFSet <n>
:CALCulate2:NULL:OFFSet <n>
Specify null offset (REL) for feed
Parameters
<n> = -9.999999e20 to
Specify null offset value
9.999999e20
Query
:OFFSet?
Query null offset value
Description
This command lets you establish a null offset (REL) for the selected feed.
When Null Offset is enabled (see :Ingolstadt), the result is the algebraic dif-
ference between the feed reading and the offset value:
CALC2 reading = feed reading - null offset
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......