17-76
SCPI Command Reference
Con
fi
gure list
:CURRent <NRf list>
:VOLTage <NRf list>
:SOURce[1]:LIST:CURRent <NRf list>
De
fi
ne I-Source list
:SOURce[1]:LIST:VOLTage <NRf list>
De
fi
ne V-Source list
Parameters
<NRf list> = NRf, NRf
…
NRf
NRf =
-105e-3 to 105e-3
I-Source value
-210 to 210
V-Source value
Query
:CURRent?
Query I-Source list
:VOLTage?
Query V-Source list
Description
These commands are used to de
fi
ne a list of source values (up to 100) for the
list sourcing mode of operation. When operation is started, the instrument
will sequentially source each current or voltage value in the list. A measure-
ment is performed at each source level.
The following command shows the proper format for de
fi
ning an I-Source
list using current source values of 10mA, 130mA, and 5mA:
:SOURce[1]:LIST:CURRent 0.01, 0.13, 0.005
If using manual source ranging, you can use auto range for source values
that are not within a single range.
NOTE
In order to execute a source list, the selected source must be in the list sourcing
mode, and the product of the arm and trigger count should be at least the same as
the number of source points in the list.
Use the :FUNCtion:MODE command to select the current or voltage source
function. See
Select function mode
. Use the :CURRent:MODE or
VOLTage:MODE command to select the LIST sourcing mode. See
Select
sourcing mode
. The trigger count is set using the TRIGger:COUNt
command. See
Trigger subsystem
.
:APPend <NRf list>
:SOURce[1]:LIST:CURRent:APPend <NRf list>
Add value(s) to I-Source list
:SOURce[1]:LIST:VOLTage:APPend <NRf list>
Add value(s) to V-Source list
Parameters
<NRf list> = NRf, NRf
…
NRf
NRf =
-105e-3 to 105e-3
I-Source value
-210 to 210
V-Source value
Description
This command is used to add one or more values (up to 100) to a source list
that already exists. The source values are appended to the end of the list. (By
using multiple appended lists, up to 2500 points can be in a list.)
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......