9-20
Sweep Operation
Table 9-4 lists the command sequence for the diode programming example.
Custom sweep commands
Table 9-5 summarizes remote commands used for custom sweep operation. See Section 17,
Configure list
, for more details on these commands.
Table 9-4
Staircase sweep programming example (diode test)
Command
Description
*RST
:SENS:FUNC:CONC OFF
:SOUR:FUNC CURR
:SENS:FUNC ‘VOLT:DC’
:SENS:VOLT:PROT 1
:SOUR:CURR:START 1E-3
:SOUR:CURR:STOP 10E-3
:SOUR:CURR:STEP 1E-3
:SOUR:CURR:MODE SWE
:SOUR:SWE:RANG AUTO
:SOUR:SWE:SPAC LIN
:TRIG:COUN 10
:SOUR:DEL 0.1
:OUTP ON
:READ?
Restore GPIB default conditions.
Turn off concurrent functions.
Current source function.
Volts sense function.
1V voltage compliance.
1mA start current.
10mA stop current.
1mA step current.
Select current sweep mode.
1
Auto source ranging.
Select linear staircase sweep.
Trigger count = # sweep points.
2
100ms source delay.
Turn on source output.
Trigger sweep, request data.
1
This command should normally be sent after START, STOP, and STEP to avoid delays caused by rebuilding
sweep when each command is sent.
2
For single sweep, trigger count should equal number of points in sweep: Points = (Stop-Start)/Step + 1. You
can use SOUR:SWE:POIN? query to read the number of points.
Table 9-5
Custom sweep commands
Command
Description
:SOURce:CURRent:MODE LIST
:SOURce:VOLTage:MODE LIST
:SOURce:LIST:CURRent <list>
:SOURce:LIST:CURRent:APPend <list>
:SOURce:LIST:CURRent:POINts?
:SOURce:LIST:VOLTage < list>
:SOURce:LIST:VOLTage:APPend <list>
:SOURce:LIST:VOLTage:POINts?
:SOURce:SWEep:RANGing <name>
Select current list (custom) sweep mode.
Select voltage list (custom) sweep mode.
Define I-source (list = I1, I2,… In).
Add I-source list value(s) (list =I1, I2,…In).
Query length of I-source list.
Define V-source list (list = V1, V2,… Vn).
Add V-source list value(s) (list =V1, V2,…Vn).
Query length of V-source list.
Select source ranging (name = BEST, AUTO, or FIXed).
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......