Digital I/O Port, Interlock, and Output Configuration
12-7
Front panel output con
fi
guration
The output is con
fi
gured from the CONFIGURE OUTPUT menu and is structured as fol-
lows. Note that bullets indicate the primary items of the sweep menu, while dashes indicate
options. Use Section 1,
Rules to navigate menus
to check and/or change operate options.
Con
fi
gure OUTPUT menu
Press CONFIG and then ON/OFF OUTPUT to display the menu, which is also shown in
Figure 12-5.
•
OFF STATE
—
Use to select the OFF state of the output. (See
Output-off states
for
details.)
-
NORMAL
—
When the OUTPUT is turned off, the V-Source is selected and set to
0V. Current compliance is set to 0.5% full scale of the present current range.
NORMAL is the default off-state.
-
ZERO
—
When the V-Source OUTPUT is turned off, the V-Source is set to 0V
and current compliance is not changed. When the I-Source OUTPUT is turned off,
the V-Source mode is selected and set to 0V. Current compliance is set to the pro-
grammed Source I value or to 0.5% full scale of the present current range, which-
ever is greater. Measurements are performed and displayed while the OUTPUT is
off.
-
GUARD
—
When OUTPUT is turned OFF, the current source is selected and set
to 0A. Voltage compliance is set to 0.5% full scale of the present voltage range.
•
AUTO OFF
—
Use to ENABLE or DISABLE auto output off. When enabled, the
OUTPUT will turn off after the measurement phase of every SDM cycle. The OUTPUT
turns back on at the beginning of the next SDM cycle. When disabled, the OUTPUT
stays on as long as the SourceMeter is operating within the trigger model (ARM annun-
ciator on). With the OUTPUT enabled, pressing the ON/OFF key will disable the OUT-
PUT and disable auto output off.
•
INTERLOCK
—
Use to ENABLE or DISABLE the interlock line of the Digital out-
put. This line is used as an interlock for a test
fi
xture. See
Safety interlock
.
CONFIG
ON/OFF
OUTPUT
INTERLOCK
ENABLE
DISABLE
OFF STATE
AUTO OFF
ENABLE
DISABLE
NORMAL
ZERO
GUARD
Figure 12-5
Output configuration
menu tree
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......