Data Store
8-7
Data store programming example
Table 8-2 summarizes the commands for basic data store operation. These commands set up
the SourceMeter as follows:
•
Reading source: raw readings.
•
Number of points: 10.
•
Acquired data: buffer readings, mean (average), and standard deviation.
NOTE
You can determine when the buffer is full by reading the appropriate status register
bit. See Section 14 for details on the status structure.
Table 8-2
Data store example
Command
Description
*RST
:SOUR:VOLT 10
:TRAC:FEED SENS
:TRAC:POIN 10
:TRAC:FEED:CONT NEXT
:TRIG:COUN 10
:OUTP ON
:INIT
:TRACE:DATA?
:CALC3:FORM MEAN
:CALC3:DATA?
:CALC3:FORM SDEV
:CALC3:DATA?
Restore GPIB defaults.
Source 10V.
Store raw readings in buffer.
Store 10 readings in buffer.
Enable buffer.
Trigger count = 10.
Turn on output.
Trigger readings.
Request raw buffer readings.
Select mean buffer statistic.
Request buffer mean data.
Select standard deviation statistic.
Request standard deviation data.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......