Basic Source-Measure Operation
NPLC cache setup
Follow the steps below to enable and use NPLC caching with a source memory sweep:
1.
Press the MENU key, select A/D-CTRL, then press ENTER.
2.
Select AUTO-ZERO, then press ENTER.
3.
Choose DISABLE, then press ENTER to disable auto zero.
4.
From the A/D CONTROLS menu, select NPLC-CACHE, then press ENTER.
5.
Select ENABLE, then press ENTER to enable NPLC caching.
6.
Use the EXIT key to back out of the main menu structure.
7.
Set up the source memory parameters, and run the source memory sweep. (See
Source
memory sweep
in Section 9.)
Typical NPLC cache test times
Typically, NPLC caching will decrease source memory sweep times by a factor of three.
Table 3-3 shows typical averaged times for a test consisting of 10 sweeps of four source mem-
ory locations with NPLC values for successive memory locations set to 10, 1, 0.1, and 0.01
respectively.
V-source protection
Use V-source protection to select the maximum voltage level the SourceMeter can output.
Available limit values include 20V, 40V, 60V, 80V, 100V, 120V, 160V, and NONE (allows volt-
age >160V). These are absolute values with 5% tolerance. The power-on default is NONE.
W
A
R
N
I
N
G
Even with the voltage protection limit set to the lowest value (20V),
NEVER touch the triax cable(s) or anything connected to the terminals of
the SourceMeter when it is on or connected to an external source. Always
assume that a hazardous voltage (>30V rms) is present when the power is
on.
To prevent damage to DUT (devices under test) or external circuitry, DO
NOT program the V-Source to levels that exceed the voltage protection
limit.
T
a
b
le 3
-
3
Typical NPLC cache test times
NPLC cache conditions
Time
Auto Zero OFF, Caching OFF
5.89s
Auto Zero ON, Caching OFF
5.89s
Auto Zero OFF, Caching ON
2.05s
Auto Zero ON, Caching ON
5.89s
Auto Zero OFF, Caching ON, Cache empty
2.5s
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......