Triggering
10-9
Output trigger speci
fi
cations
The SourceMeter can be programmed to output a trigger after various trigger model actions.
See
Trigger model
. The output trigger provides a TTL-compatible output pulse that can be used
to trigger other instruments. The specifications for this trigger pulse are shown in Figure 10-4.
A trigger link line can source 1mA and sink up to 50mA.
External triggering example
In a simple test system, you may want to close a switching channel and then measure the
resistance of the DUT connected to that channel. This test system is shown in Figure 10-5,
which uses a SourceMeter to measure 10 DUTs switched by a Model 7011 multiplexer card in
a Model 7001/7002 Switch System.
TTL High
(3.4V Typical)
Meter Complete
TTL Low
(0.25V Typical)
10
µ
s
Minimum
Figure 10-4
Trigger link output pulse
specifications
Triax
Cable
HI
LO
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
Connect to REMOTE
PreAmp connector on rear
panel of mainframe
1
DUT
#1
2
DUT
#2
10
DUT
#10
OUTPUT
Card 1
7011 MUX Card
Remote
PreAmp
Preamp
Cable
IN/OUT
HIGH
Figure 10-5
DUT test system
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......