Bante210 Benchtop pH Meter 

Instruction Manual 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
BANTE INSTRUMENTS LIMITED 

 

 

Summary of Contents for Bante210

Page 1: ...Bante210 Benchtop pH Meter Instruction Manual BANTE INSTRUMENTS LIMITED...

Page 2: ...Adapter pH Electrode Temperature Probe pH Buffer Pouches DC9V Power Adapter Display Bante210 benchtop pH meter is equipped with a clear and bright LCD display that used to show measured values mode i...

Page 3: ...er the calibration mode Press and hold the key to enter the setup menu o C Press the key to set the temperature or enter the temperature calibration mode Press the key to increase setting value Press...

Page 4: ...over from the electrode arm 2 Use the screwdriver to tighten the screw moderately 3 Insert the plastic cover to previous position Installation is completed Connectors The meter provides 4 connectors f...

Page 5: ...on the sensor cord Always make sure that the connector is clean and dry Connecting the Temperature Probe Bante210 benchtop pH meter comes with a temperature probe for measuring temperature of the sam...

Page 6: ...ng the pH Buffer Solutions Bante210 benchtop pH meter is packaged with three pH buffer packets required for calibration Open the pH7 00 buffer packet place the powder into a 250ml volumetric flask Pou...

Page 7: ...ss any key for 3 hours the meter will turn off automatically To disable the auto off function please read the SETUP MENU section Setup Menu Bante210 benchtop pH meter contains an integrated setup menu...

Page 8: ...to confirm The meter shows HOLD NO indicating that the auto hold function is disabled If the auto hold function is enabled the meter will automatically sense a stable endpoint reading and freeze it If...

Page 9: ...ould not be reset WARNING Reset function will restore the meter back to factory default parameters all calibration values and selected parameters will be reset 7 Press or key to enable or disable the...

Page 10: ...or The meter is now switched to automatic temperature compensation mode MANUAL TEMPERATURE COMPENSATION 1 DO NOT connect the temperature probe to meter Press o C key to enter the temperature setting m...

Page 11: ...INGLE POINT CALIBRATION 1 1 Make sure that you have selected 1 point calibration in the setup menu 1 2 Rinse the pH electrode with distilled water 1 3 Press CAL key the meter shows CAL1 pH7 00 or CAL1...

Page 12: ...p menu 3 2 Repeat steps 1 2 to 1 5 above When the first calibration point is done the display will show CAL2 pH4 01 The meter prompts you to continue with second point calibration 3 3 Rinse the pH ele...

Page 13: ...n If the electrode slope or calibration result does not meet measurement requirements the indicator will disappear on the display Temperature Calibration During the measurement when automatic temperat...

Page 14: ...inse the pH electrode in distilled water store the sensor into the electrode storage solution CLEANING THE ELECTRODE Salt deposits Soak the pH electrode in warm tap water to dissolve deposits then tho...

Page 15: ...t Cosmetics Dairy Products Education Fats Cream Field Use Fish Products Lab Flasks Low Ionic Meat Cheese Micro Samples Paint Photographic Soil Surface Test Tubes Tris Buffer Viscose Samples Troublesho...

Page 16: ...t Calibration Ranges Measured value 10 C Others Temperature Compensation 0 100 C 32 212 F Manual or Automatic Hold Function Manual or Automatic Power Off Manual or Automatic 180 minutes after last key...

Page 17: ...w to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover sensor and calibration solutions Out of warranty products will be...