SCPI Command Reference 17-103
Abort source/measure cycle
:ABORt
Abort operation
Description
When this action command is sent, the SourceMeter aborts operation and
returns to the idle state.
A faster way to return to idle is to use the DCL or SDC command.
With auto output-off enabled (:SOURce1:CLEar:AUTO ON), the output
will remain on if operation is terminated before the output has a chance to
automatically turn off.
Program trigger model
:COUNt <n>
:ARM[:SEQuence[1]][LAYer[1]]:COUNt <n>
Set arm count
:TRIGger[:SEQuence[1]]:COUNt <n>
Set trigger count
Parameters
<n> =
1 to 2500
Specify count (see NOTE)
DEFault
Sets count to 1
MINimum
Sets count to 1
MAXimum
See NOTE
INFinite
(ARM:COUNt only)
NOTE
The product of arm count and trigger count cannot exceed 2500.
Query
:COUNt?
Queries programmed count
:COUNt? DEFault
Queries *RST default count
:COUNt? MINimum
Queries lowest allowable count
:COUNt? MAXimum
Queries largest allowable count
Description
This command is used to specify how many times an operation is performed
in the speci
fi
ed layer of the trigger model.
For example, assume the arm count is set to 2 and the trigger counter is set
to 10, the SourceMeter is con
fi
gured to perform 10 source-measure opera-
tions twice for a total of 20 source-measure operations.
The product of the arm count and trigger count cannot exceed 2500. If, for
example, the arm count is 2, then the maximum trigger count is 1250.
NOTE
INFinite can be used only with ARM:COUNt, and FETCh?, READ?, MEAS?,
CALC1:DATA?, or CALC2:DATA? cannot be used with infinite arm count. Only
INIT will start measurements, and only interlock, over-temperature, SDC, DCL, or
ABORt should be used to stop the sweep.
ARM:COUNt INFinite can be used for repetitive source waveforms or for
long tests where only the last reading is important. For example, the limits
could be used to drive the interlock to abort a test when some condition is
met. DATA? would then give the answer to the test.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......