EOT MODE
—
Use this menu item to control the operation of Digital I/O line 4 to act
as an EOT (End of Test) or BUSY signal:
-
EOT
—
In 3-bit mode, automatically output a HI pulse on Digital I/O line 4 at end
of test. In 4-bit mode, EOT is not automatically controlled.
-
/EOT
—
In 3-bit mode, automatically output a LO pulse on Digital I/O line 4 at
end of test. In 4-bit mode, this option is not available.
-
BUSY
—
Set Digital I/O line 4 HI while unit is busy. With BUSY selected, the
unit behaves as if it is in 3-bit mode.
-
/BUSY
—
Set Digital I/O line 4 LO while unit is busy. With /BUSY selected, the
unit behaves as if it is in 3-bit mode.
Performing limit tests
Perform the following steps to run limit tests:
Step 1: Con
fi
gure test system.
As previously explained in Section 2, your test system could be as simple as connecting a
DUT to the SourceMeter or could employ the use of a handler for binning operations. Adding a
scanner to the test system allows you to test multi-element devices (such as resistor networks).
Make sure that the Digital I/O is con
fi
gured appropriately for the handler you are using.
Step 2: Con
fi
gure source-measure functions.
Con
fi
gure the SourceMeter for the desired source-measure operations as follows:
1.
Select the desired source function by pressing SOURCE V or SOURCE I.
2.
Set the source level and compliance limit to the desired values.
CONFIG
LIMIT
DIGOUT
H/W
LIMITS
MODE
SIZE
FAIL
MODE
CONTROL
AUTO
CLEAR
S/W
LIMITS
PASS
EOT
MODE
LIM2,3
LIM5-12
CONTROL
HILIM
LOLIM
SRC MEM
LOC
DIG I/O
PATTERN
/EOT
EOT
/BUSY
BUSY
Figure 11-10
Limits configuration
menu tree
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......