Limit Testing
11-19
Remote limit testing
Limit commands
Table 11-1 summarizes remote commands to control limit testing. See
CALCulate2
and
SOURce2
in Section 17 for more details on these commands.
T
a
b
le 11
-
1
Limit commands
Command*
Description*
:CALCulate2:FEED <name>
Select limit test input path (name = CALCulate[1],
VOLTage, CURRent, or RESistance).
:CALCulate2:DATA?
Acquire limit test data.
:CALCulate2:LIMit:COMPliance:FAIL <name>
Set Limit 1 fail condition. Name = IN (fail into
compliance) or OUT (fail out of compliance).
:CALCulate2:LIMitX:LOWer <n>
Specify lower Limit X; X = 2, 3, 5-12 (n = limit).
:CALCulate2:LIMitX:UPPer <n>
Specify upper Limit X; X = 2, 3, 5-`2 2 (n = limit).
:CALCulate2:LIMit[1]:COMPliance:SOURce2 <NRf> |
<NDN>
Specify limit 1 fail bit pattern.
(NRf | NDN = pattern).
:CALCulate2:LIMitX:LOWer:SOURce2 <NRf> | <NDN> Specify lower Limit X fail bit pattern for grading
mode; X = 2, 3, 5-12 (NRf | NDN = bit pattern).
:CALCulate2:LIMitX:UPPer:SOURce2 <NRf> | <NDN>
Specify upper Limit X fail bit pattern for grading
mode; X = 2, 3, 5-12 (NRf = bit pattern).
:CALCulate2:LIMitX:PASS:SOURce2 <NRf> | <NDN>
Specify pass pattern for sorting mode
(NRf | NDN = bit pattern).
:CALCulate2:LIMit[1]:STATe <state>
Enable/disable Limit 1 test (state = ON or OFF).
:CALCulate2:LIMitX:STATe <state>
Enable/disable Limit X test; X = 2, 3, 5-12
(state = ON or OFF).
:CALCulate2:LIMit[1]:FAIL?
Query Limit 1 test result (0 = pass, 1 = fail).
:CALCulate2:LIMitX:FAIL?
Query Limit X test result; X = 2, 3, 5-12
(0 = pass, 1 = fail).
:CALCulate2:CLIMits:PASS:SOURce2 <NRf> | <NDN>
Specify pass bit pattern (NRf | NDN = pattern).
Sorting mode only if limits 2, 3 and 5-12
disabled.
:CALCulate2:CLIMits:FAIL:SOURce2 <NRf> | <NDN>
Specify fail bit pattern for sorting mode
(NRf | NDN = pattern).
:CALCulate2:CLIMits:PASS:SMLocation <location>
Specify pass source memory location. Location =
NRf (memory #) or NEXT (next location).
:CALCulate2:CLIMits:FAIL:SMLocation <location>
Specify fail source memory location. (Location =
NRf (memory #) or NEXT (next location).
:CALCulate2:CLIMits:BCONtrol <name>
Control I/O port pass/fail update. Name =
IMMediate (at failure) or END (end of sweep).
:CALCulate2:CLIMits:CLEar
Clear test results, reset I/O port.
:CALCulate2:CLIMits:CLEar:AUTO <state>
Enable/disable auto-clear (state = ON or OFF).
:CALCulate2:CLIMits:MODE <name>
Select Digital I/O control mode (name = GRADing
or SORTing).
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......