Instruction Manual (Operation)

COMPACT WATER QUALITY METER
LAQUAtwin-Na-11

Specifications

Items in package

Consumable parts sold separately

Part Names

Note

Press the switches 0.5 seconds or more unless
otherwise specified.

Initial Setup

Attaching/detaching the sensor

Note

Be careful not to twist the waterproof gasket.

Inserting/removing batteries

Electrode conditioning

Note

Before using the sensor for the first time or
after several days of disuse, perform electrode
conditioning. 

Perform calibration after electrode condition-
ing.

1. Place some drops of 2000 ppm standard

solution to the flat sensor.

2. Wait a few hours before use.

There is no need to switch the meter ON.

3. Clean the flat sensor with running water.

Model

LAQUAtwin-Na-11

Target

Sodium ion (Na

+

)

Measurement principle

ISE method

Minimum sample volume

More than 0.3 mL

*1

Measure-
ment range

2 to 9900 [unit: ppm or mg/L]

Resolution 
(default)

Display range: Resolution

2 to 99:

1

100 to 990:

10

1000 to 9900: 100 

[unit: ppm or mg/L]

Calibration

Up to 2 points 
Default: 150 ppm and 2000 ppm

Accuracy

*2

±10% of reading value

Waterproof

IP67 (no failure when immersed in 
water at a depth of 1 m for 30 min) 

*3

Display

Custom (monochrome) digital LCD 
with backlight

Operating 
environment

Temperature: 5

C to 40

C

Humidity: 85% relative humidity or 
less (no condensation)

Power

CR2032 batteries (

2)

Battery life

Approx. 400 h continuous operation 
(backlight off mode)

*4

Material

ABS epoxy (main material)

Dimensions 164 

 29 

 20 mm (excluding projec-

tions)

Mass

Approx. 50 g (excluding batteries)

*1 0.05 mL or more if sampling sheet B (sold sepa-

rately) is used.

*2 The closeness of agreement between a mea-

sured value and an actual value of standard solu-
tion after two-point calibration using 150 ppm and 
2000 ppm standard solutions.

The standard solution used for the later cali-
bration was measured after two-point calibra-
tion at 150 ppm and 2000 ppm. 

The calibration and measurement are per-
formed at the same temperature.

The error of standard solutions and rounding
error (

1 digit) are not included.

*3 The meter cannot be used underwater.
*4 When the backlight is used, battery life will 

shorten.

Items

Quantity

Sensor

S022

1

Meter

1

Storage case

1

Batteries

CR2032

2

Standard solution

150 ppm

1

2000 ppm

1

Pipette

1

Sampling sheet B (5 sheet-pack)

1

Instruction manual (Operation)

1

Instruction manual (Before use)

1

Items

Specifications

Part No.

Sensor

S022, Na

+

 

3200459867

Standard 
solution

Y022L, Na

+

 150 ppm

3200457724

Y022H, Na

+

 2000 ppm

3200457723

Sampling 
sheet B

Y046, 100 sheet-pack

3200053858

1

Flat sensor

2

Light shield cover

3

Liquid junction

4

Response mem-
brane

5

Lithium batteries

6

MEAS switch

7

ON/OFF switch

8

CAL switch

9

Waterproof gasket

10 Strap eyelet

Attaching the sensor

1. Power OFF the meter.
2. 
Confirm that the waterproofing gasket is

clean and undamaged.

3. Slide the sensor

onto the meter so
that catch "A" on the
back of the meter fits
into hole "a" on the
sensor tongue as
shown.

1

2

3

4

5

6

7 10

8

9

a

A

Waterproof
gasket

Detaching the sensor

1. Power OFF the meter.
2. 
Lift the sensor tongue

tip and slide the sensor
a little away from the
meter. 

3. Pull out the sensor all the way from the

meter.

Inserting the batteries

1. Power OFF the meter.
2. 
Slide both batteries into

the battery case as
shown.

Be sure to use two
CR2032 batteries, and put
them with the plus sides
(+) upwards.

Removing the batteries

1. Power OFF the meter.
2. 
Use a ball-point pen or

other tool to pry the batter-
ies out from the clips as
shown.

Battery

Battery clip

Summary of Contents for LAQUAtwin-Na-11

Page 1: ... The closeness of agreement between a mea sured value and an actual value of standard solu tion after two point calibration using 150 ppm and 2000 ppm standard solutions The standard solution used for the later cali bration was measured after two point calibra tion at 150 ppm and 2000 ppm The calibration and measurement are per formed at the same temperature The error of standard solutions and rou...

Page 2: ...asurement mode automatically 5 Open the light shield cover and remove the standard solution Then remove moisture on the sensor by gently dabbing with a soft tissue This completes the 1st point calibration 6 To perform 2nd point calibration repeat steps 2 to 5 Calibration error If blinks and Er4 error dis play appears the calibration has failed Perform electrode conditioning Check that the correct ...

Page 3: ...perature sensor adjustment To perform accurate measurement with correc tion for temperature effects follow the steps below Normally this is not necessary 1 Ready a reference thermometer and allow the meter and reference thermometer to reach to room temperature Sampling sheet holder cover Part No 3200459736 Light shield cover Sampling sheet B Part No 3200053858 2 Set the display mode to temperature...

Page 4: ...same time Er4 is dis played during the calibra tion Please note that if you press the CAL switch in mV or temperature display mode Er4 is displayed This is because there is no calibration facility available for these modes Er1 is dis played soon power ON The internal IC in the meter may be defective Perform meter initializa tion If Er1 is still displayed after the initial ization the internal IC i...

Page 5: ... concen tration is changed the calibration data are initialized 2nd calibration point setting The concentration of the 2nd calibration point can be set In calibration point setting pressing the MEAS switch increases the displayed value After the displayed value reaches 9800 it returns to 3 All segment indication 2 s Software ver indication 1 s ON OFF switch MEAS switch CAL switch ppm default mg L ...