17
CO2 with CAPNOSTAT Mainstream,
CapnoFlex LF, and Dual CO2 modules
CO
2
compatibility limitations
CAPNOSTAT Mainstream, CapnoFlex LF and Dual CO2 modules with the B850 only.
For detailed information regarding module, monitor, and accessory compatibility, see
the CARESCAPE Modular Monitors Supplemental Information Manual.
CO₂ safety precautions
CO
2
warnings
WARNING
Always inspect the airway adapter for a tight connection and
proper operation before attaching it to the patient.
WARNING
Leaks in the gas sampling circuit (water trap and sampling
line) may cause inaccurate readings.
WARNING
Remove the airway sampling line from the patient’s airway
while nebulized medications are being delivered.
WARNING
Handle the water trap and its contents as you would any body
fluid. Infectious hazard may be present.
WARNING
Since sample gas may contain anesthetic agents, make sure
that it is not released in the room. Connect exhaust to a
scavenging system to prevent exposure to anesthetic agents.
WARNING
Route all tubing away from the patient’s throat to avoid
strangulation.
WARNING
To avoid the spread of infectious disease, do not allow the
exhaust to discharge in the direction of the patient or user.
WARNING
CapnoFlex LF or CAPNOSTAT modules: O
2
, N
2
O and anesthetic
agent gases may interfere with EtCO
2
readings.
2062971-001
CARESCAPE Modular Monitors
307
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...