Messages
Message
Location
Explanation
What to do
●
Calibrating
●
param.
Calibration is in progress.
●
No action required.
●
Calibration fail
●
param.
Unsuccessful calibration.
●
Authorized service personnel
should repeat the calibration
procedure.
●
CCI high
/
CCI low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
CCO high
/
CCO low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Check Ext Device
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
C.O. Complete
●
C.O.
menu
Successful cardiac output
determination has been
performed, but the module
is not yet ready for a new
measurement. Also when the 15
minute timeout for confirming
the C.O. measurement expires
with the
Cardiac Output
>
Measurement
menu open.
●
Wait until the message
disappears.
●
CO measurement
removed
●
al. area
Acquisition module has been
removed.
●
Connect the module if
you want to restart the
measurement.
●
C.O. out of range
●
C.O.
menu
The measurement results are
invalid.
●
Perform a new measurement.
●
Confirm C.O.
●
al. area,
param.
Measurement data has not
been confirmed before trying
to exit the
Cardiac Output
>
Measurement
menu.
If more than 15 minutes have
passed since the start of the C.O.
measurement, this message
resets automatically.
●
Select
Confirm C.O.
●
Curve over range
●
C.O.
menu
Thermodilution waveform values
are over the valid range.
●
Wait for the blood
temperature to stabilize
and retry the measurement.
●
Curve under range
●
C.O.
menu
Thermodilution waveform values
are under the valid range.
●
Wait for the blood
temperature to stabilize
and retry the measurement.
●
Identical C.O.
modules
●
al. area
There are two or more identical
C.O. modules in the system.
Also when there are more than
one active C.O. sources.
●
Remove all but one C.O.
module.
2062971-001
CARESCAPE Modular Monitors
483
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...