EEG and auditory evoked potentials
The numbers and letters of the 10-20 system refer to electrode locations:
●
Odd numbered electrodes: placed on the left
●
Even numbered electrodes: placed on the right
●
Letters: F = frontal; T = temporal; C = central; P = parietal; O = occipital; Z = midline
electrodes
Connecting the EEG leadset
1.
Connect the headbox to the module.
2.
Select a suitable preconfigured leadset or your own montage.
3.
Connect the leadset to the headbox.
4.
Attach the electrodes to the patient as indicated in the leadset or according to
your own montage.
5.
Observe the results of the electrode check. Reconnect the electrodes if the
impedance is too high.
Attaching EEG electrodes within hair area
Preferably use cup electrodes if montage includes placements within hair area.
1.
Mark the spots on the patient’s head according to the montage you have chosen.
2.
Comb or cut the hair away from the spots and rub the skin with abrasive paste to
remove oil and grease.
3.
Clean the skin with alcohol.
4.
Attach the electrodes using conductive paste.
NOTE
Use water to detach the electrodes.
Attaching EEG electrodes outside hair area
Preferably use adhesive electrodes if all montage placements are within skin area.
1.
Clean the skin with alcohol.
2.
Remove the electrode from its package and attach it to the desired location.
3.
Press the electrode gently for several seconds to attach it.
376
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...