Messages
Message
Location
Explanation
What to do
●
No ext device
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
Over range
●
param.
Measured FiO
2
is more than
103%.
●
Calibrate airway gases.
●
Over scale
●
wavef.
Gas signal exceeds the
maximum waveform area.
●
Check the patient status.
●
Select a larger scale for
waveform.
●
Purging
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
Replace Water Trap
●
al. area,
param.
Water trap is partially blocked.
●
Replace the water trap.
●
RR (CO2) High
/
RR
(CO2) Low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Sample line blocked
●
al. area
The gas sampling line is blocked
or the water trap is occluded.
●
Change sampling line and
water trap.
●
Sample gas out
●
param.
The water trap is not connected,
the sample gas outflow is
blocked, or there is a leak inside
the module.
●
Check water trap connection.
●
Remove the blockage from
the sample gas outlet.
●
Change module if needed.
●
Service CO2 module
●
param.
Technical failure in the module.
●
Contact authorized service
personnel.
●
Service gas module
●
Service gas module
- and specific error
indication
●
al.area,
param.,
wavef.
The measuring sensor is
inoperative or the temperature
in the module has increased.
●
Contact authorized service
personnel.
●
Service ext device
●
param.
Interfaced device should be
checked.
●
Contact authorized service
personnel.
●
Service Mode
●
param.
Interfaced device should be
checked.
●
Contact authorized service
personnel.
●
Standby
●
param.
Acquisition module has shut
down the pump.
●
No action required.
●
Warming up
●
param.
The sensor is warming up.
●
Wait until the message
disappears.
●
Zero error
●
param.
Zeroing has failed.
●
Repeat the zeroing.
●
Zeroing
●
param.
Zeroing is in progress.
●
Wait until the zeroing is
completed.
492
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...