Messages
Message
Location
Explanation
What to do
●
SvO2 high
/
SvO2 low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
SvO2 measurement
removed
●
al. area
Acquisition module has been
removed.
●
Connect the module if
you want to restart the
measurement.
●
SvO2 not calibrated
●
al. area,
param.
The optical module is connected
to the monitor and the catheter
has not been calibrated.
●
Perform in vivo calibration.
●
SvO2 out of range
●
param.
Values are above or below the
range the monitor can process.
●
Check cables and
connections.
●
Repeat in vivo calibration.
●
Change the optical module.
●
If the catheter is faulty,
replace it.
●
If the cable or module is
broken, contact authorized
service personnel.
●
SvO2 signal poor
●
al. area,
param.
There is signal pulsation, the
catheter is touching the wall,
or there is an intensity shift in
signal quality level.
●
Flush the catheter.
●
Check the optical module and
connections.
●
Check the catheter
placement.
●
SvO2 temp error
●
al. area
The temperature of the optical
module is out of range for more
than 10 minutes.
●
Check the optical module and
connections.
●
Replace the optical module.
●
Contact authorized service
personnel.
●
Wait, initializing
●
param.
Optical module signal is being
processed.
●
Wait until the message
disappears.
●
Warming up
●
param.
The optical module is warming
up.
●
Wait until the module has
warmed up and the message
disappears.
Messages related to gases measurement
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
488
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...