Messages
Message
Location
Explanation
What to do
●
NIBP cuff occlusion
●
al. area
Occlusion during measurement
or overpressured cuff.
●
Check the cuff.
●
NIBP cuff loose
●
al. area
Loose cuff or cuff hose.
●
Check the cuff and cuff hose.
●
NIBP Dia high
/
NIBP
Dia low
●
NIBP Mean high
/
NIBP Mean low
●
NIBP Sys high
/
NIBP
Sys low
●
al. area,
param.
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
NIBP manual
●
al. area
Loose cuff or cuff hose.
●
Check the cuff and cuff hose.
●
NIBP measurement
removed
●
al. area
Acquisition module has lost NIBP
communication.
●
Remove the module and then
reconnect it.
●
Replace the module.
●
If the problem persists,
contact authorized service
personnel.
●
Protect calibration
●
param.
Calibration is not protected.
●
Contact authorized service
personnel.
●
Select cuff size
●
al. area,
param.
Cuff does not have an
automatically detectable
cuff ID.
●
Select a cuff size from the
NIBP setup menu.
●
Select inflation limits
●
al. area,
param.
Cuff does not have an
automatically detectable
cuff ID.
●
Select an inflation limit from
the NIBP setup menu.
●
Unable to meas. Dia
●
param.
Accurate diastolic pressure is
difficult to measure because of
artifacts, weak pulsation etc.
●
Assess the patient and check
inflation limits.
●
Perform a new measurement.
●
Unable to meas. Sys
●
param.
Systolic pressure probably
higher than maximum inflation
pressure or artifacts interfere in
the systolic area.
●
Assess the patient and check
the cuff placement.
●
Perform a new measurement.
●
Unstable zero press.
●
param.
Pressure is unstable at start of
the NIBP measurement.
●
Check the patient status.
●
Check hose and cuff position.
●
Repeat the measurement.
●
If the problem persists,
contact authorized service
personnel.
●
Weak pulsation
●
param.
Weak or unstable oscillation
signal.
●
Check the patient status.
●
Reposition the cuff.
●
Repeat the measurement.
474
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...